Title: Dosimetric Consistency of Telecobalt and Brachytherapy Installations over a period of thirteen years: An Institutional Study

Authors: Mithu Barthakur MSc., DRP, Bhaveshwar Yadav MSc., Shachindra Goswami MSc., DRP, Pranjal Goswami MSc., DRP, Shashi Bhushan Sharma MSc., Gautam Sarma MD, Apurba.Kr Kalita MD

 DOI: https://dx.doi.org/10.18535/jmscr/v8i8.41

Abstract

Introduction: The use of radionuclides in telecobalt and brachytherapy machines demands safe handling of the equipment and accurate measurement of source activity to ensure safety of the patients, workers and public. This paper reports dosimetric consistency of few Quality Assurance parameters of Theratron Phoenix Telecobalt and micro Selectron HDR Brachytherapy machines performed for a period of thirteen years from 2007 to 2019.

Materials and Method: The output of the telecobalt unit for both SSD and SAD techniques and the source strength of the HDR Brachytherapy unit in terms of RAKR were measured as per national and international guidelines. The percentage variations between measured and theoretically calculated values were obtained for the thirteen years period. Timer error, error in Timer linearity and variations in TPS vs. manually calculated treatment times were calculated for the HDR unit after each source replacements.

Results: The results obtained from the measurements done all over the thirteen years period were found to be well within the stipulated values specified by competent authority. The results were shown in graphical and tabulated form.

Conclusion: The results show that the parameters measured were well within tolerance limits for a long period of thirteen years. Hence, this study establishes the consistency and accuracy in the dosimetric parameters of the Theratron Phoenix and micro Selectron HDR Brachytherapy machines.

Keywords:  Telecobalt, Brachytherapy, Reference Air Kerma Rate, Timer Error and Linearity, Dosimetric Consistency, TPS.

Editorial Policy

Authors should prepare their manuscripts according to the instructions given in the authors' guidelines. Manuscripts which do not ..

Read More.....

Frequency of Publication

JMSCR is published as monthly journal with 12 issues per year. Special editions are also planned subjected to the scope and need....

Read more...

Submission of Articles

Authors are invited to submit their research articles, review papers, Case Report properly formatted as per the author guidelines.........

Read more...