Title: Study on Platelet Count on the Basis of Red Cell: Platelet Ratio

Authors: Dr Ajamal Singh Bhayal, Dr Mahendra Kumar Gupta, Dr Narendra Kumar Sharma, Dr Asim Singh, Mr Manoj Kumar Yadav

 DOI:  http://dx.doi.org/10.18535/jmscr/v3i10.08

Abstract

The modern hematology analyzers are able to produce platelet counts with great precision and accuracy. Our aim to verification of the reliability of the estimation technique of platelet count on the basis of red cell: platelet ratio. An overnight fast venous blood samples was collected for the following parameters: Manual Method & Automated Method. The results of platelet count using the manual method were as follows: the range was between 100-499 x103/µl, the mean platelet count was 301.46±44.49 x 103/µl and by using the automated method, platelet count ranged between 95-484 x103/µl, the mean was 328.47±54.00 x103/µl. Red blood cell: platelet ratio method requires only an accurate RBC count performed on a calibrated hematology analyzer to calculate platelet count. This method is precise, simple and consumes less time than using a counting chamber and therefore, potentially should supersede ordinary manual counting.

KEY WORD:  Red cell: platelet ratio, Manual Method & Automated Method

Editorial Policy

Authors should prepare their manuscripts according to the instructions given in the authors' guidelines. Manuscripts which do not ..

Read More.....

Frequency of Publication

JMSCR is published as monthly journal with 12 issues per year. Special editions are also planned subjected to the scope and need....

Read more...

Submission of Articles

Authors are invited to submit their research articles, review papers, Case Report properly formatted as per the author guidelines.........

Read more...